Use of PIDs for Soil Headspace Measurements

quipnotes3.gif (4724 bytes)

RAE Systems Technical Note

EnviroEquip News, May 2002

Placing the probe of an valuable PID in contaminated, moist soil isn't the most ideal situation.  However by following a few simple steps you can keep your readings accurate and your PID safe from damage.  These notes, RAE Systems Technical Note TN-118, provide useful tips on using RAE PIDs for soil headspace measurements.

Measuring organic compounds emitted from potentially contaminated soils requires special attention beyond that needed for typical ambient air monitoring.  Soils are often dusty and humid, conditions that can cause high, drifting readings on MiniRAE PIDs if not properly maintained.  Interferences are usually traceable to condensation in the sensor, causing a current leakage across the electrodes and thus a false positive signal.  The situation is exacerbated when the sensor is contaminated by soil dust or condensed, high boiling organic compounds.

Preparing a Sample

Place the sample into a clean sample container or bag.  Ensure that enough air space ("headspace") above the sample is present for sampling.  Shake the sample container to thoroughly mix the soil sample with the air in the headspace.  Let the sample equilibrate to room temperature of approximately 25 deg C.  Sample through a septum in a hard container or through the sample bag wall.  A septum can easily be formed in virtually any hard container with aluminium foil held in place by a rubber band around the rim of the vessel.  When sampling with the PID, ensure that dirt and moisture are not sucked into the PID probe.

To ensure optimum performance:

1) Keep the sensor clean using high-purity methanol, preferably using an ultrasonic bath.  

2) Keep the lamp clean using high- purity methanol.  Never use acetone.

3) Use the C-filter to absorb moisture and dust.  Perform frequent changes of the C-filter and membrane filters (daily to weekly depending on usage and dirtiness).

4) Use the water trap filters as an extra precaution, especially in dusty or moist environments where water mist may be present.

5) Avoid situations in which the PID is colder than the soil being sampled, such as heating the soil samples to increase headspace organic concentration, or bringing a cold PID into a warm room without allowing time for temperature equilibrium.  If anything, try to keep the PID warmer than the soil samples.

6) To obtain more stable readings, plumb the effluent flow from the MiniRAE PID back into the sample container to reduce the losses.  Use Teflon or metal tubing for this purpose so as to prevent adsorption to Tygon or other plastic tubing.  Losses will not be stopped altogether but will be greatly reduced.

MiniRAE 2000 PID

 

EnviroEquip Product Links

EnviroEquip QuipNotes

 

 

 

_ QUIPNOTES INDEX _
 

A division of Biolab (Aust) Pty Ltd
Privacy | Disclaimer | Trademarks | Contact

Copyright Biolab (Aust) Pty Ltd 2007